Bard Peripheral Vascular recall: EnCor EnCompass Breast Biopsy and Tissue Removal System Console
Bard Peripheral Vascular Inc is recalling EnCor EnCompass Breast Biopsy and Tissue Removal System Console – 99. The U.S. Food and Drug Administration published the recall on October 7, 2026; it began on August 20, 2026. Distributed nationwide.
Serious Class II means the product may cause temporary or medically reversible health problems, or the chance of serious harm is remote.

Share
| Product | EnCor EnCompass Breast Biopsy and Tissue Removal System Console |
| Brand | Bard Peripheral Vascular |
| Company | Bard Peripheral Vascular Inc |
| Type | Medical devices |
| Classification | Class II |
| Recall date | August 20, 2026 |
| Published by FDA | October 7, 2026 |
| Quantity | 99 |
| Recall ID | 99590 |
Recalled products
- EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with software
- FDA recall number
- Z-3275-2026
- Codes, lots and dates
- Software Versions 1.1.1.10207 and 1.2.0.12133. UDI-DI: 00801741235221 and 00801741234248. Serial Numbers: VTLNPX06, VTLNPX0F, VTLNPX0D, VTLNPX08, VTLNPX0K, VTLNPX0A, VTLNPX0C, VTLNPX0L, VTLNPX0J, VTLNPX04, VTLNPX0H, VTLNPX02, VTLPPX1G, VTLPPX1D, VTLPPX1C, VTLPPX1M, VTLPPX2H, VTLPPX2I, VTLPPX2F, VTLQPX15, VTLQPX1Q, VTLQPX1G, VTLQPX14, VTLQPX0T, VTLQPX1Y, VTLQPX1M, VTLQPX2G, VTLQPX1O, VTLQPX1K, VTLQPX19, VTLQPX2E, VTLQPX1W, VTLQPX10, VTLQPX1Z, VTLQPX24, VTLQPX21, VTLQPX0X, VTLQPX1E, VTLQPX0U, VTLQPX26, VTLQPX2F, VTLQPX25, VTLQPX0Z, VTLQPX18, VTLQPX28, VTLQPX20, VTLQPX27, VTLQPX11, VTLQPX2B, VTLQPX1F, VTLQPX2H, VTLQPX41, VTLQPX3T, VTLQPX43, VTLRPX0R, VTKYPX09, VTKYPX44, VTKYPX46, VTKYPX47, VTKYPX48, VTKYPX49, VTKZPX14, VTKZPX15, VTKZPX20, VTKZPX17, VTLNPX4I, VTLNPX49, VTLNPX4G, VTLNPX4N, VTLNPX4E, VTLNPX4L, VTLNPX44, VTLNPX46, VTLNPX4C, VTLPPX27, VTLPPX28, VTLPPX29, VTLPPX1Z, VTLPPX37, VTLPPX3B, VTLPPX34, VTLPPX38, VTLPPX36, VTLQPX0Q, VTLQPX0B, VTLQPX0E, VTLQPX0P, VTLQPX0F, VTLRPX0X, VTLRPX0Z, VTKYPX07, VTKYPX06, VTKYPX04, VTLNPX4B, VTLNPX42, VTLNPX4F, VTLPPX3D, VTLQPX0M, VTKYPX51
- Quantity
- 99
- Classification
- Class II
Identifiers and details
- FDA event ID
- 99590
- Initiated
- Voluntary: Firm initiated
- Status
- Ongoing
- Firm location
- Tempe, AZ, United States
Why it was recalled
The reason the FDA lists: “Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.”
Where it was sold
Nationwide. Distribution as the FDA records it: “US Nationwide distribution in the states of KS, SC, CA, TX, GA, NC, PA, SC, ID, MI, KY, CT, IL, VA, MA, WI, NY, IN, TN, OH, UT.”
What to do
Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.
Official notice: FDA Enforcement Report record (event 99590, openFDA) – the same recall numbers can be searched in the FDA Enforcement Report.
Other medical device recalls
| Date | Recall | Problem | Agency |
|---|---|---|---|
| Oct 7, 2026 | Potential cracks in the hose, which could lead to leakage or the hose falling off. Most serious | FDA Class I | |
| Oct 7, 2026 | a fire hazard Most serious | FDA Class I | |
| Oct 7, 2026 | Due to incorrect inside packaging label for eyepiece. Serious | FDA Class II | |
| Oct 7, 2026 | Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxy… Serious | FDA Class II | |
| Oct 7, 2026 | The potential for falsely elevated results. Serious | FDA Class II | |
| Oct 7, 2026 | A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a sof… Serious | FDA Class II |
About this recall
Is Bard Peripheral Vascular EnCor EnCompass Breast Biopsy recalled?
Yes. Bard Peripheral Vascular Inc recalled it on August 20, 2026; FDA published it October 7, 2026 (Class II). Only the lots, dates or models listed on this page are included.
Why was it recalled?
Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.
What should I do if I have it?
Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.
Where was it sold?
Nationwide.
More recalls
Source: recall records published by the U.S. Food and Drug Administration (openFDA Enforcement Reports), the USDA Food Safety and Inspection Service, the Consumer Product Safety Commission and the National Highway Traffic Safety Administration, collected twice a day. Product names, lot codes, dates and model numbers are shown exactly as the agency lists them; the agency's notice is the authority. 77.today is not affiliated with any of these agencies.