FDA · Class II · August 26, 2026

Beckman Coulter recall: DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1)… and 1 more product

Beckman Coulter Inc. is recalling Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1)… and 1 more product. The U.S. Food and Drug Administration published the recall on August 26, 2026; it began on July 15, 2026. Distributed nationwide.

Serious Class II means the product may cause temporary or medically reversible health problems, or the chance of serious harm is remote.

An FDA laboratory that tests medical devices
Illustration: An FDA laboratory that tests medical devices – not the recalled product. Photo: U.S. Food and Drug Administration, public domain.
ProductBeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1)… and 1 more product
BrandBeckman Coulter
CompanyBeckman Coulter Inc.
TypeMedical devices
ClassificationClass II
Recall dateJuly 15, 2026
Published by FDAAugust 26, 2026
Recall ID99436
As listed by FDA

Recalled products (2)

  1. Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers
    FDA recall number
    Z-2973-2026
    Codes, lots and dates
    1) Part Number C63522, UDI/DI 14987666545089; 2) Part Number C63521, UDI/DI 14987666545072; 3) Part Number C13252, UDI/DI 15099590742331.
    Classification
    Class II
  2. Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
    FDA recall number
    Z-2974-2026
    Codes, lots and dates
    1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
    Classification
    Class II
Exactly as the agency lists them

Identifiers and details

FDA event ID
99436
Initiated
Voluntary: Firm initiated
Status
Ongoing
Firm location
Brea, CA, United States

Why it was recalled

The reason the FDA lists: “a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.”

Where it was sold

Nationwide. Distribution as the FDA records it: “Worldwide distribution - US Nationwide.”

What to do

Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.

Official notice: FDA Enforcement Report record (event 99436, openFDA) – the same recall numbers can be searched in the FDA Enforcement Report.

Latest

Other medical device recalls

DateRecallProblemAgency
Oct 7, 2026Draeger recall: VentStar Resus NeoMedical devices · 1,063,643 unitsPotential cracks in the hose, which could lead to leakage or the hose falling off.
Most serious
FDA
Class I
Oct 7, 2026Medline Industries recall: Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories and 1 more productMedical devices · 771,127 totala fire hazard
Most serious
FDA
Class I
Oct 7, 2026Karl Storz Endoscopy recall: HOPKINS Telescope 30¿Medical devices · 6 eyepiecesDue to incorrect inside packaging label for eyepiece.
Serious
FDA
Class II
Oct 7, 2026Noxbox recall: NOxBOXi Nitric Oxide Delivery SystemMedical devices · 1,667 unitsPotential for irregular oxygen readings due to a manufacturing issue affecting certain oxy…
Serious
FDA
Class II
Oct 7, 2026Siemens Healthcare Diagnostics recall: Atellica CH Revised C-Reactive ProteinMedical devices · 42940 units (3938 US, 39002 OUS)The potential for falsely elevated results.
Serious
FDA
Class II
Oct 7, 2026Tandem Diabetes Care recall: T:slim Mobile ApplicationMedical devices · 177A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a sof…
Serious
FDA
Class II
FAQ

About this recall

Is Beckman Coulter Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1)… recalled?

Yes. Beckman Coulter Inc. recalled it on July 15, 2026; FDA published it August 26, 2026 (Class II). Only the lots, dates or models listed on this page are included.

Why was it recalled?

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

What should I do if I have it?

Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.

Where was it sold?

Nationwide.

Recalls

More recalls

Source: recall records published by the U.S. Food and Drug Administration (openFDA Enforcement Reports), the USDA Food Safety and Inspection Service, the Consumer Product Safety Commission and the National Highway Traffic Safety Administration, collected twice a day. Product names, lot codes, dates and model numbers are shown exactly as the agency lists them; the agency's notice is the authority. 77.today is not affiliated with any of these agencies.

Cookies

We measure traffic so we know which stories you care about. You can change your choice at any time.