FDA · Class II · September 2, 2026

GE Medical Systems Information Technologies recall: Portrait Core Services Software V1.1 and 3 more products

GE Medical Systems Information Technologies Inc is recalling Portrait Core Services Software V1.1 and 3 more products – 23 units; 4 units; 15 units; 0. The U.S. Food and Drug Administration published the recall on September 2, 2026; it began on July 20, 2026. Distributed in 2 states (Michigan, New York).

Serious Class II means the product may cause temporary or medically reversible health problems, or the chance of serious harm is remote.

An FDA laboratory that tests medical devices
Illustration: An FDA laboratory that tests medical devices – not the recalled product. Photo: U.S. Food and Drug Administration, public domain.
ProductPortrait Core Services Software V1.1 and 3 more products
BrandGE Medical Systems Information Technologies
CompanyGE Medical Systems Information Technologies Inc
TypeMedical devices
ClassificationClass II
Recall dateJuly 20, 2026
Published by FDASeptember 2, 2026
Quantity23 units; 4 units; 15 units; 0
Recall ID99536
As listed by FDA

Recalled products (4)

  1. PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209
    FDA recall number
    Z-3036-2026
    Codes, lots and dates
    GTIN 00195278679284, Serial Numbers: SWT25250001TA, SWT25360001TA, SWT25360002TA, SWT24090006TA, SWT25100001TA, SWT25120002TA, SWT24140007TA, SWT24150008TA, SWT25100004TA, SWT24050001TA, SWT25110003TA, SWT24180011TA, SWT25120003TA, SWT25230001TA, SWT25230002TA, SWT25110009TA, SWT25110010TA, SWT25230003TA, SWT25230005TA, SWT25230006TA, SWT25230004TA, SWT25260001TA, SWT31240012TA.
    Quantity
    23 units
    Classification
    Class II
  2. PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592
    FDA recall number
    Z-3037-2026
    Codes, lots and dates
    GTIN 00198953056752, Serial Numbers: SWT25510001TA, SWT26110010TA, SWT26070001TA, SWT26120008TA.
    Quantity
    4 units
    Classification
    Class II
  3. PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353
    FDA recall number
    Z-3038-2026
    Codes, lots and dates
    GTIN 00195278364838, Serial Numbers: SUC23080050TA, SUD22470001TA, SUD22420001TA, SUD22470010TA, SUD23390001TA, SUD23390003TA, SUD23390004TA, SUD23390009TA, SUD23390010TA, SUD22470009TA, SUD22470003TA, SUD22470004TA, SUD22470005TA, SUD23390005TA, SUD22280002TA.
    Quantity
    15 units
    Classification
    Class II
  4. FRU (Field replaceable unit), PORTRAIT CORE SERVICES SW RELOAD V1.1, Ref# 5700225
    FDA recall number
    Z-3039-2026
    Codes, lots and dates
    UDI-DI (GTIN) 00195278679307
    Quantity
    0
    Classification
    Class II
Exactly as the agency lists them

Identifiers and details

FDA event ID
99536
Initiated
Voluntary: Firm initiated
Status
Ongoing
Firm location
Waukesha, WI, United States

Why it was recalled

The reason the FDA lists: “Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.”

Where it was sold

2 states: Michigan, New York. Distribution as the FDA records it: “US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.”

What to do

Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.

Official notice: FDA Enforcement Report record (event 99536, openFDA) – the same recall numbers can be searched in the FDA Enforcement Report.

Same brand

Other GE Medical Systems Information Technologies recalls

DateRecallProblemAgency
Sep 23, 2026GE Medical Systems Information Technologies recall: Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04Medical devices · 29,570 (23,632 US; 5,938 OUS)Device may not resume patient monitoring for at least 3 minutes following system restart. …
Serious
FDA
Class II
Latest

Other medical device recalls

DateRecallProblemAgency
Oct 7, 2026Draeger recall: VentStar Resus NeoMedical devices · 1,063,643 unitsPotential cracks in the hose, which could lead to leakage or the hose falling off.
Most serious
FDA
Class I
Oct 7, 2026Medline Industries recall: Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories and 1 more productMedical devices · 771,127 totala fire hazard
Most serious
FDA
Class I
Oct 7, 2026Karl Storz Endoscopy recall: HOPKINS Telescope 30¿Medical devices · 6 eyepiecesDue to incorrect inside packaging label for eyepiece.
Serious
FDA
Class II
Oct 7, 2026Noxbox recall: NOxBOXi Nitric Oxide Delivery SystemMedical devices · 1,667 unitsPotential for irregular oxygen readings due to a manufacturing issue affecting certain oxy…
Serious
FDA
Class II
Oct 7, 2026Siemens Healthcare Diagnostics recall: Atellica CH Revised C-Reactive ProteinMedical devices · 42940 units (3938 US, 39002 OUS)The potential for falsely elevated results.
Serious
FDA
Class II
Oct 7, 2026Tandem Diabetes Care recall: T:slim Mobile ApplicationMedical devices · 177A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a sof…
Serious
FDA
Class II
FAQ

About this recall

Is GE Medical Systems Information Technologies Portrait Core Services Software V1.1 recalled?

Yes. GE Medical Systems Information Technologies Inc recalled it on July 20, 2026; FDA published it September 2, 2026 (Class II). Only the lots, dates or models listed on this page are included.

Why was it recalled?

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

What should I do if I have it?

Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.

Where was it sold?

In 2 states (Michigan, New York).

Recalls

More recalls

Source: recall records published by the U.S. Food and Drug Administration (openFDA Enforcement Reports), the USDA Food Safety and Inspection Service, the Consumer Product Safety Commission and the National Highway Traffic Safety Administration, collected twice a day. Product names, lot codes, dates and model numbers are shown exactly as the agency lists them; the agency's notice is the authority. 77.today is not affiliated with any of these agencies.

Cookies

We measure traffic so we know which stories you care about. You can change your choice at any time.