Medline Industries recall: Medline Micro-Kill Bleach Germicidal Bleach Wipes
Medline Industries, LP is recalling Medline Micro-Kill Bleach Germicidal Bleach Wipes – 591,561 units; 447,618 units; 1,244 units; 31,728. The U.S. Food and Drug Administration published the recall on August 19, 2026; it began on June 30, 2026. Distributed nationwide.
Serious Class II means the product may cause temporary or medically reversible health problems, or the chance of serious harm is remote.

Share
| Product | Medline Micro-Kill Bleach Germicidal Bleach Wipes |
| Brand | Medline Industries |
| Company | Medline Industries, LP |
| Type | Medical devices |
| Classification | Class II |
| Recall date | June 30, 2026 |
| Published by FDA | August 19, 2026 |
| Quantity | 591,561 units; 447,618 units; 1,244 units; 31,728 |
| Recall ID | 99432 |
Recalled products (4)
- Medline Micro-Kill Bleach Germicidal Bleach Wipes,7" x 8", 150 count, 150 count, Product Number MSC351400AN
- FDA recall number
- Z-2923-2026
- Codes, lots and dates
- UDI-DI (each) 10888277652514, UDI-DI (case) 20888277652511, Lot Numbers: All lots produced between 05/20/2025 and 05/8/2026
- Quantity
- 591,561 units
- Classification
- Class II
- Medline Micro-Kill Bleach Germicidal Bleach Wipes, 6" x 5", 150 count, Product Number MSC351410AN
- FDA recall number
- Z-2924-2026
- Codes, lots and dates
- UDI-DI (each) 10888277652521, UDI-DI (case) 20888277652528, Lot Numbers: All lots produced between 05/13/2025 and 04/21/2026
- Quantity
- 447,618 units
- Classification
- Class II
- Medline Micro-Kill Bleach Germicidal Bleach Wipes, 12" x 12", 110 Count, Product Number MSC351450AN
- FDA recall number
- Z-2925-2026
- Codes, lots and dates
- UDI-DI (each) 80080196314521, UDI-DI (case) 10080196314522, Lot Numbers: All lots produced between 10/27/2025 and 11/19/2025
- Quantity
- 1,244 units
- Classification
- Class II
- Medline Micro-Kill Bleach Germicidal Bleach Wipes, 12" x 12", 110 Count Bag, Product Number MSC351450BG4
- FDA recall number
- Z-2926-2026
- Codes, lots and dates
- UDI-DI (each) 00193489015242, UDI-DI (case) 20193489015246, Lot Numbers: All lots produced between 05/27/2025 and 04/10/2026
- Quantity
- 31,728
- Classification
- Class II
Identifiers and details
- FDA event ID
- 99432
- Initiated
- Voluntary: Firm initiated
- Status
- Ongoing
- Firm location
- Northfield, IL, United States
Why it was recalled
The reason the FDA lists: “Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.”
Where it was sold
Nationwide. Distribution as the FDA records it: “US Nationwide distribution in the states and territories of Guam, Puerto Rico.”
What to do
Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.
Official notice: FDA Enforcement Report record (event 99432, openFDA) – the same recall numbers can be searched in the FDA Enforcement Report.
Other Medline Industries recalls
| Date | Recall | Problem | Agency |
|---|---|---|---|
| Oct 7, 2026 | a fire hazard Most serious | FDA Class I | |
| Sep 30, 2026 | The kits contain neurological sponges with the potential for unexpected variability of end… Serious | FDA Class II | |
| Sep 30, 2026 | Kits contain Healthmark Instrument Channel Brushes, which were recalled by the manufacture… Serious | FDA Class II | |
| Sep 23, 2026 | The kits contain certain lots of sterile applicators which may contain fungal contaminatio… Most serious | FDA Class I | |
| Sep 23, 2026 | a sterility problem Serious | FDA Class II | |
| Sep 23, 2026 | a sterility problem Serious | FDA Class II | |
| Sep 9, 2026 | Convenience kits contain BD ChloraPrep Clear - 1mL Applicators from specific lots that wer… Serious | FDA Class II | |
| Sep 9, 2026 | a sterility problem Serious | FDA Class II | |
| Sep 2, 2026 | Component labeling discrepancy Serious | FDA Class II | |
| Sep 2, 2026 | Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neuros… Serious | FDA Class II | |
| Sep 2, 2026 | Neonatal and infant heated wire breathing circuits may experience sporadic failure. The he… Serious | FDA Class II | |
| Sep 2, 2026 | Kits contain a potentially contaminated component which were recalled by the manufacturer Serious | FDA Class II |
All 13 Medline Industries recalls →
About this recall
Is Medline Industries Medline Micro-Kill Bleach Germicidal Bleach Wipes recalled?
Yes. Medline Industries, LP recalled it on June 30, 2026; FDA published it August 19, 2026 (Class II). Only the lots, dates or models listed on this page are included.
Why was it recalled?
Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.
What should I do if I have it?
Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.
Where was it sold?
Nationwide.
More recalls
Source: recall records published by the U.S. Food and Drug Administration (openFDA Enforcement Reports), the USDA Food Safety and Inspection Service, the Consumer Product Safety Commission and the National Highway Traffic Safety Administration, collected twice a day. Product names, lot codes, dates and model numbers are shown exactly as the agency lists them; the agency's notice is the authority. 77.today is not affiliated with any of these agencies.