FDA · Class II · October 7, 2026

Shanghai United Imaging Healthcare recall: Computed Tomography X-ray System and 1 more product

Shanghai United Imaging Healthcare Co., Ltd. is recalling Computed Tomography X-ray System and 1 more product – 4; 23; 1; 14; 11; 53; 34; 7; 3; 6. The U.S. Food and Drug Administration published the recall on October 7, 2026; it began on August 11, 2026. Distributed nationwide.

Serious Class II means the product may cause temporary or medically reversible health problems, or the chance of serious harm is remote.

An FDA laboratory that tests medical devices
Illustration: An FDA laboratory that tests medical devices – not the recalled product. Photo: U.S. Food and Drug Administration, public domain.
ProductComputed Tomography X-ray System and 1 more product
BrandShanghai United Imaging Healthcare
CompanyShanghai United Imaging Healthcare Co., Ltd.
TypeMedical devices
ClassificationClass II
Recall dateAugust 11, 2026
Published by FDAOctober 7, 2026
Quantity4; 23; 1; 14; 11; 53; 34; 7; 3; 6
Recall ID99731
As listed by FDA

Recalled products (12)

  1. Computed Tomography X-ray System, Model: uCT 530 with software
    FDA recall number
    Z-3277-2026
    Codes, lots and dates
    UDI-DI: 06971576831012. Serial Numbers/Software: 300174/R001.19.5.1839204-Re3300SP1H1_P4, 304006/R001.17.6.1607336-Re3100SP4, 304007/R001.19.4.1810621-Re3300SP1H1_P3, 304008/R001.19.4.1810621-Re3300SP1H1_P3
    Quantity
    4
    Classification
    Class II
  2. Computed Tomography X-ray System, Model: uCT 550 with software
    FDA recall number
    Z-3278-2026
    Codes, lots and dates
    UDI-DI: 06971576831036. Serial Numbers/Software: 353021/R001.19.6.1857994-Re3300SP1H1_P5, 353045/R001.19.6.1857994-Re3300SP1H1_P5, 353049/R001.19.6.1857994-Re3300SP1H1_P5, 354001/R001.19.6.1857994-Re3300SP1H1_P5, 354004/R001.19.6.1857994-Re3300SP1H1_P5, 354005/R001.19.6.1857994-Re3300SP1H1_P5, 354007/R001.19.6.1857994-Re3300SP1H1_P5, 354009/R001.19.6.1857994-Re3300SP1H1_P5, 354013/R001.19.6.1857994-Re3300SP1H1_P5, 354015/R001.19.6.1857994-Re3300SP1H1_P5, 354016/R001.19.6.1857994-Re3300SP1H1_P5, 354018/R001.17.6.1607336-Re3100SP4, 354020/R001.17.6.1607336-Re3100SP4, 354021/R001.17.6.1607336-Re3100SP4, 354022/R001.19.4.1810621-Re3300SP1H1_P3, 354023/R001.19.6.1857994-Re3300SP1H1_P5, 354024/R001.19.6.1857994-Re3300SP1H1_P5, 354025/R001.19.6.1857994-Re3300SP1H1_P5, 354026/R001.19.6.1857994-Re3300SP1H1_P5, 354027/R001.19.6.1857994-Re3300SP1H1_P5, 354028/R001.19.6.1857994-Re3300SP1H1_P5, 354029/R001.19.6.1857994-Re3300SP1H1_P5, 354030/R001.19.6.1857994-Re3300SP1H1_P5
    Quantity
    23
    Classification
    Class II
  3. Computed Tomography X-ray System, Model: uCT 760 with software
    FDA recall number
    Z-3279-2026
    Codes, lots and dates
    UDI-DI: 06971576831043. Serial Number/Software: 600271/R003.8.4.1810621-Re3300SP1H1_P3
    Quantity
    1
    Classification
    Class II
  4. Computed Tomography X-ray System, Model: uCT 780 with software
    FDA recall number
    Z-3280-2026
    Codes, lots and dates
    UDI-DI: 06971576831050. Serial Number/Software: 684001/R001.17.6.1607336-Re3100SP4, 684002/R001.17.6.1607336-Re3100SP4, 684003/R003.8.4.1810621-Re3300SP1H1_P3, 684004/R003.8.4.1810621-Re3300SP1H1_P3, 684005/R003.8.4.1810621-Re3300SP1H1_P3, 684006/R003.8.4.1810621-Re3300SP1H1_P3, 684007/R003.8.4.1810621-Re3300SP1H1_P3, 684008/R003.8.4.1810621-Re3300SP1H1_P3, 684009/R003.8.4.1810621-Re3300SP1H1_P3, 684010/R003.8.4.1810621-Re3300SP1H1_P3, 684011/R003.8.4.1810621-Re3300SP1H1_P3, 684012/R003.9.0.1776134-Re-3300SP2, 684013/R003.11.0.1897081-Re3310SP1, 684014/R003.11.0.1897081-Re3310SP1
    Quantity
    14
    Classification
    Class II
  5. Computed Tomography X-ray System, Model: uCT ATLAS with software
    FDA recall number
    Z-3281-2026
    Codes, lots and dates
    UDI-DI: 06971576831104. Serial Number/Software: 863001/R001.17.6.1607336-Re3100SP4, 863002/R001.17.6.1607336-Re3100SP4, 863004/R001.17.6.1607336-Re3100SP4, 863005/R001.17.6.1607336-Re3100SP4, 863007/R001.21.0.1897081-Re3310SP1, 863008/R001.21.0.1897081-Re3310SP1, 869001/R001.22.0.2012850-Re3320, 869002/R001.19.4.1810621-Re3300SP1H1_P3, 869003/R001.21.0.1897081-Re3310SP1, 869004/R001.21.0.1897081-Re3310SP1, 11CT863010/R001.17.6.1607336-Re3100SP4
    Quantity
    11
    Classification
    Class II
  6. Computed Tomography X-ray System, Model: uCT ATLAS Astound with software
    FDA recall number
    Z-3282-2026
    Codes, lots and dates
    UDI-DI: 06971576831340. Serial Number/Software: 803011/R002.9.0.1897081-Re3310SP1, 803012/R002.4.6.1607336-Re3100SP4, 803013/R002.4.6.1607336-Re3100SP4, 803014/R002.9.0.1897081-Re3310SP1, 803016/R002.4.6.1607336-Re3100SP4, 803017/R002.9.0.1897081-Re3310SP1, 803018/R002.9.0.1897081-Re3310SP1, 803019/R002.4.6.1607336-Re3100SP4, 803020/R002.9.0.1897081-Re3310SP1, 803022/R002.9.0.1897081-Re3310SP1, 803023/R002.9.0.1897081-Re3310SP1, 803024/R002.9.0.1897081-Re3310SP1, 809001/R002.8.1.1840891-Re3310A_P1, 809002/R002.9.0.1897081-Re3310SP1, 809003/R002.8.1.1840891-Re3310A_P1, 809004/R002.4.6.1607336-Re3100SP4, 809005/R002.4.6.1607336-Re3100SP4, 809006/R002.8.1.1840891-Re3310A_P1, 809007/R002.4.7.1671762-Re3100SP4_P1, 809008/R002.4.7.1671762-Re3100SP4_P1, 809009/R002.4.7.1671762-Re3100SP4_P1, 809010/R002.4.7.1671762-Re3100SP4_P1, 809011/R002.4.7.1671762-Re3100SP4_P1, 809012/R002.4.7.1671762-Re3100SP4_P1, 809013/R002.4.7.1671762-Re3100SP4_P1, 809014/R002.7.4.1810621-Re3300SP1H1_P3, 809015/R002.7.4.1810621-Re3300SP1H1_P3, 809016/R002.9.0.1897081-Re3310SP1, 809017/R002.9.0.1897081-Re3310SP1, 809018/R002.7.4.1810621-Re3300SP1H1_P3, 809019/R002.7.4.1810621-Re3300SP1H1_P3, 809020/R002.9.0.1897081-Re3310SP1, 809021/R002.7.4.1810621-Re3300SP1H1_P3, 809022/R002.7.4.1810621-Re3300SP1H1_P3, 809023/R002.7.4.1810621-Re3300SP1H1_P3, 809024/R002.7.4.1810621-Re3300SP1H1_P3, 809025/R002.7.4.1810621-Re3300SP1H1_P3, 809026/R002.7.4.1810621-Re3300SP1H1_P3, 809027/R002.7.4.1810621-Re3300SP1H1_P3, 809028/R002.9.0.1897081-Re3310SP1, 809029/R002.7.4.1810621-Re3300SP1H1_P3, 809030/R002.7.4.1810621-Re3300SP1H1_P3, 809031/R002.7.4.1810621-Re3300SP1H1_P3, 809032/R002.7.4.1810621-Re3300SP1H1_P3, 809033/R002.9.0.1897081-Re3310SP1, 809034/R002.9.0.1897081-Re3310SP1, 809035/R002.9.0.1897081-Re3310SP1, 809036/R002.9.0.1897081-Re3310SP1, 809037/R002.9.0.1897081-Re3310SP1, 809038/R002.9.0.1897081-Re3310SP1, 809039/R002.9.0.1897081-Re3310SP1, 809040/R002.9.0.1897081-Re3310SP1, 809041/R002.9.0.1897081-Re3310SP1
    Quantity
    53
    Classification
    Class II
  7. Positron Emission Tomography and Computed Tomography System, Model: uMI 550 with software
    FDA recall number
    Z-3283-2026
    Codes, lots and dates
    UDI-DI: 06971576832026. Serial Number/Software: 200045/R001.19.0.1812568, 200163/R001.19.3.2011750, 200164/R001.18.0.1683822, 200165/R001.18.0.1683822, 200166/R001.18.0.1683822, 200167/R001.18.0.1683822, 202009/R001.19.3.2011750, 202016/R001.19.0.1812568, 202018/R001.19.0.1812568, 202019/R001.19.0.1812568, 202020/R001.19.1.1929746, 202021/R001.19.1.1929746, 202022/R001.19.1.1929746, 230040/R001.19.3.2011750, 230048/R001.19.0.1812568, 230106/R001.19.0.1812568, 232038/R001.19.0.1812568, 232039/R001.19.0.1812568, 232040/R001.19.0.1812568, 232041/R001.19.0.1812568, 232042/R001.19.0.1812568, 232043/R001.19.0.1812568, 232044/R001.19.0.1812568, 232045/R001.19.0.1812568, 232046/R001.19.0.1812568, 232047/R001.19.0.1812568, 232048/R001.19.1.1929746, 232049/R001.19.1.1929746, 232050/R001.19.1.1929746, 232051/R001.19.1.1929746, 202023/R001.19.1.1929746, 202024/R001.19.1.1929746, 232052/R001.19.1.1929746, 232053/R001.19.1.1929746
    Quantity
    34
    Classification
    Class II
  8. Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama 35 with software
    FDA recall number
    Z-3284-2026
    Codes, lots and dates
    UDI-DI: 06971576832170. Serial Number: 11MI410003/R001.19.4.1909524, 11MI410004/R001.19.4.1909524, 11MI410005/R001.19.4.1909524, 20MI410001/R001.19.4.1909524, 20MI410002/R001.19.4.1909524, 20MI410003/R001.21.0.1993334, 20MI410004/R001.19.4.1909524
    Quantity
    7
    Classification
    Class II
  9. Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama GS with software
    FDA recall number
    Z-3285-2026
    Codes, lots and dates
    UDI-DI: 06971576832187. Serial Number: 11MIP00007/R001.17.5.1785268, 11MIP00009/R001.18.2.1967222, 11MIP00010/R001.18.2.1967222
    Quantity
    3
    Classification
    Class II
  10. Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo with software
    FDA recall number
    Z-3286-2026
    Codes, lots and dates
    UDI-DI: 06971576832231. Serial Number/Software: 11MI500004/R002.4.0.1872339, 11MI500005/R002.4.0.1872339, 11MI500006/R002.4.0.1872339
    Quantity
    3
    Classification
    Class II
  11. Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo S with software
    FDA recall number
    Z-3287-2026
    Codes, lots and dates
    UDI-DI: 06971576832255. Serial Number/Software: 11MI510005/R003.2.0.1872339, 20MI510001/R003.2.0.1872339, 20MI510002/R003.2.0.1872339, 20MI510003/R003.2.0.1872339, 20MI510004/R003.2.0.1872339, 20MI510005/R003.2.0.1872339
    Quantity
    6
    Classification
    Class II
  12. Positron Emission Tomography and Computed Tomography System, Model: uMI 780
    FDA recall number
    Z-3288-2026
    Codes, lots and dates
    UDI-DI: 06971576832033. Serial Number/Software: 300098/R001. 19.0.1812568
    Quantity
    1
    Classification
    Class II
Exactly as the agency lists them

Identifiers and details

FDA event ID
99731
Initiated
Voluntary: Firm initiated
Status
Ongoing
Firm location
Shanghai, China

Why it was recalled

The reason the FDA lists: “Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.”

Where it was sold

Nationwide. Distribution as the FDA records it: “US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.”

What to do

Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.

Official notice: FDA Enforcement Report record (event 99731, openFDA) – the same recall numbers can be searched in the FDA Enforcement Report.

Latest

Other medical device recalls

DateRecallProblemAgency
Oct 7, 2026Draeger recall: VentStar Resus NeoMedical devices · 1,063,643 unitsPotential cracks in the hose, which could lead to leakage or the hose falling off.
Most serious
FDA
Class I
Oct 7, 2026Medline Industries recall: Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories and 1 more productMedical devices · 771,127 totala fire hazard
Most serious
FDA
Class I
Oct 7, 2026Karl Storz Endoscopy recall: HOPKINS Telescope 30¿Medical devices · 6 eyepiecesDue to incorrect inside packaging label for eyepiece.
Serious
FDA
Class II
Oct 7, 2026Noxbox recall: NOxBOXi Nitric Oxide Delivery SystemMedical devices · 1,667 unitsPotential for irregular oxygen readings due to a manufacturing issue affecting certain oxy…
Serious
FDA
Class II
Oct 7, 2026Siemens Healthcare Diagnostics recall: Atellica CH Revised C-Reactive ProteinMedical devices · 42940 units (3938 US, 39002 OUS)The potential for falsely elevated results.
Serious
FDA
Class II
Oct 7, 2026Tandem Diabetes Care recall: T:slim Mobile ApplicationMedical devices · 177A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a sof…
Serious
FDA
Class II
FAQ

About this recall

Is Shanghai United Imaging Healthcare Computed Tomography X-ray System recalled?

Yes. Shanghai United Imaging Healthcare Co., Ltd. recalled it on August 11, 2026; FDA published it October 7, 2026 (Class II). Only the lots, dates or models listed on this page are included.

Why was it recalled?

Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.

What should I do if I have it?

Compare the recall numbers, lot numbers and expiration dates with what you have. Do not stop a prescribed medicine or device without talking to your pharmacist or doctor – they can arrange a replacement from an unaffected lot.

Where was it sold?

Nationwide.

Recalls

More recalls

Source: recall records published by the U.S. Food and Drug Administration (openFDA Enforcement Reports), the USDA Food Safety and Inspection Service, the Consumer Product Safety Commission and the National Highway Traffic Safety Administration, collected twice a day. Product names, lot codes, dates and model numbers are shown exactly as the agency lists them; the agency's notice is the authority. 77.today is not affiliated with any of these agencies.